After initialization display terminal displays in 0.0000 g in normal weighing mode. Record the observations in the balance calibration record. DOCX PDF: 6: Weighing By Transposition 1986. Established in 2007 by METTLER TOLEDO, GWP is a standardized methodology for the secure selection, calibration and operation of weighing equipment. Accuracy: – The accuracy of the balance is checked by weighing at least three different certified weights that cover the usual weighing range of the balance. The weighing procedure can be separated into three basic steps: planning, checking the balance, and weighing the material. • Record weight measurements, maintenance and corrective action in a logbook • Have the Class 1 weights re-certified annually or before certificate expires • Have balance serviced by a professional yearly 3. marized procedure for verification of scales compatibility to the process is shown in Figure 1 and outlined below: 1. 3.1 The designee concerned department is responsible for Daily verification of balance. 5.7.2 Note the weight . (2) Zero and span the balance within 12 h before weighing any filter. 2. nach ISO 9000ff, TS 16949, VDA 6.1, FDA, GLP, GMP, ...). For the latest updates pertaining to this compilation visit the relevant website. 10.2 Check the balance or scale to ensure that it is 6.0 Daily verification of weighing balance: 6.1 Verify the balance Daily before the start of the weighing, also verify the balance after a power failure, relocation and whenever it is required. edge and be measured at each location in the following sequence: center, front left, back left, back right, front right, and (optional) again center. 4.9 Ensure that there is no weight on the platform of the balance when it is not in use. 3. 5.3 Check for zero error and correct if any by pressing Zero button. 6.4 Switch on the main power supply of the balance. Allow the balance for 30 min for stabilization. Adjust level, if not leveled. Related Topic: weighing balance Calibration. Eccentricity:- The eccentricity test should be carried out using a weight equivalent to at least 30% of the maximum capacity of the balance (or of the maximum weight used on the balance). Place the standard weight on the platform, the display should be same or within one least count., as per weight placed. 5.5 Following remedies should be done to eliminate zero error in case of Daily verification of weighing balance: 5.6 Record the weight displayed and remove the material from the platform. Verification:- Verification of the balance is performed by placing a suitable weight (depending on the type of balance) in the center of the weighing pan once and comparing the result with pre-defined acceptance criteria. 4.2 Ensure that the balance platform/ pen is not touching to any surface. Related: Drift and Its Importance in Analytical Balances 5.7 For precision 5.7.1 Place 5 gm. Then, two sub-sets of weights are made and their masses are recorded. 3. Verify that the balance, scale, or load cell is in good operating condition with sufficiently small process standard deviation as verified by a valid control chart or preliminary experiments to ascertain its performance quality. The drift can also be extracted from the trend analysis of the verification test. Place the empty receiver (In case of material is highly volatile, add some amount of diluent) on the balance in the center of the pan and press the appropriate tare (Zero) key on the balance (On stabilization of indicator). 4.1 Avoid impact /mechanical shocks to the balance. Entry and Exit Procedure for staff, visitors in Pharmaceuticals, Pharmaceutical Sampling, types, tools ( Guidelines), Capsules, Types, and formulation (Complete guide), Procedure for Destruction microbial waste by autoclaving. It is recommended. Reproduction and Distribution of the same without written permission is prohibited. Keep the balance in a vibration-free environment. They should not be exposed to direct sunlight since it can cause temperature variations inside the weighing chamber. PRECAUTION: Use the trolley balance for material weighing within its verification range only. This process should be used to verify scale accuracy and reinforce quality importance. 5.2 Operation procedure 5.2.1 To switch the balance on briefly touch the “On/Off” key after the main power supply of balance has been switched on. 4.3 Before operation ensures that balance is clean, dry and labeled properly. Suggested process for verifying analytical balance accuracy and performance 1. 6.2.2 Place the standard weights near the instrument. This section describes three verifications. The weight should be placed between halfway to ¾ of the distance from the center of the pan to its. 4. 6.1 General. Press zero adjustment ‘⇒ 0 ⇐ ‘ on terminal before starting weighing process. Result will be note down the reading in the balance verification record. Preparations before calibration. maximum weight used on the balance), depending on the type of balance. 5.2 Check that the spirit level of the balance is in its proper position. If using an electronic balance, allow to warm up for at least 60 minutes. 1.2.2. 4.5 check the spirit level of the balance for its proper position. 1.2 Many laboratory balances have more than one range. Annexure for daily verification of balance: Save my name, email, and website in this browser for the next time I comment. 5.5 Following remedies should be done to eliminate zero error in case of Daily verification of weighing balance: check the balance using a spirit level. (Proposed criterion: RSD not more than 0.05%, calculated from all weighings at different locations on the pan). 5. © PharmaState Blog and pharmastate.com | Pharmaceutical Industry Guidelines, SOPs, News, Views & Articles by Industry Experts. If the display is not matching to the weight placed and the difference is more than one least count than open the side pot and adjust with the screwdriver till it matches. Repeat the procedure for five times and record the readings. The external check is done for three factors: 1. % – percentageg – Gramkg – KilogramLC – Least countNLT – Note Less ThanSOP – Standard Operating procedurewt. Calibration Weights and Calibration Mass Sets with Calibration certification and Nist Traceable Calibration of laboratory balances. 3.2 Head Production/ engineering/ warehouse /Quality Assurance and quality control are responsible to ensure overall compliance with this is SOP. Our primary manufacturers of mass and test weights include Rice Lake Weighing Systems, Sartorius, and Troemner. View all posts by PharmaState Blog. The calibration covers a range starting with a 10 mg weight, and the provider will not go any lower (I believe thay don't have the weights to do so). 5.6.5 Record the reading in Annexure-II. Applications with mandatory verification of balances and test weights include: - in commercial trade when the price of a commodity is determined by weighing Check the balance leveling gauge to make sure bubble is inside the target. Verification:-Verification of the balance is performed by placing a suitable weight (depending on the type of balance) in the center of the weighing pan once and comparing the result with pre-defined acceptance criteria. 6.5 Tolerance: The variation (if any) should be ± least count of the balance or ± 0.2 % of the certified value of standard weight used whenever is higher and for analytical balance, the variation should be ± least count of the balance or ± 0.1 % of the certified value of standard weight used whichever is higher. Our lab is using an analytical balance and it is calibrated annually by an external provider. You have to record ten measurements with the 10 mg weight … Before you can start the calibration of the weighing instrument, you … Weighing balance Calibration in Pharmaceutical, SOP on Operation, cleaning of Leak test apparatus, SOP for Camera vision system and pinhole detector, Accelerated stability testing (study) Important Questions. Using a 3-1 Weighing Design (IR5672) May 2019 Job Aids: SOP 5 Calibration-DWright V18 EXCEL | SOP 5-Control Chart DWright V03 EXCEL. Perform accuracy check of weigh balances once a year, approximately six months after vendor calibration, using reference weights. 4. (Proposed criterion: tolerance 0.5%, k=3 for class I balances and 1%, k=3 for class II and III balances). 1. (1) Independent verification of PM balance performance within 370 days before weighing any filter. Touch “Adjust.int” icon on the screen to enable internal adjustment. 5.7 Switch off the balance and main power supply at the end of day. This is not just a simple or basic procedure, wherein we will just compare a reference standard in one set up or position then finished (compared to verification only). Remove the material, which is touching the balance and check for zero error. Precision:- The precision of the balance should be verified by weighing at least 5 times a weight that is equivalent to approximately 50% of the maximum capacity of the balance. Procedure 10.1 All data are to be recorded on the "Balance or ScaleCalibration" form asshown onfigure 1. 6.3 Check the level of the balance with the help of spirit level. How to examine the repeatability of the balance in routine testing? c. The same weight should always be used in these verifications. Check SOP (or new process requirements) for weighing setpoints and limits. It must be remembered, however, that taring should only be carried out initially when performing either the single point check or repeatability user checks. It is recommended that the weights have approximately 5%, 50% and 100% of the maximum capacity of the balance (or of the. (Proposed criterion: SD = max 5*d, where d = (actual) scale interval (e.g. if we require adjust the level using leveling screw which are provided at the base of the balance. Record the observations in balance calibration record format. 6.2 Operate the balance as described in step 5.1 To 5.7. The correlation coefficient is calculated by comparing the nominal and measured masses of the weights. We do not guarantee the accuracy of this information or any results and the user is advised to take any decision or action on his/her sole discretion & responsibility. 6.4 Record the displayed weight in respective verification formats. Dust off the balance pan with a clean soft brush. This procedure can be used in your internal calibration laboratory to calibrate most of your digital weighing scale. It is determined from technical data and the external calibration certificate. The same weight should always be used in these verifications. Drift: Normally, a 10mg standard weight is used to carry out a drift check. All compilations/publications are for informational purposes only & we do not suggest to take any action/decision on the basis of this information. Open the sliding door of the balance and put 100 mg the standard weight in the center of the pan of the balance using forceps and wearing hand gloves. 5.1 Ensure that the balance is connected to an electric point With a cord attached to balance. 2. Each range will have a different accuracy, readability, and sensitivity. The following soft keys should be showing: 1. After placing the weights, verify the use range of the balance. The linearity error of the balance is an absolute value calculated by the difference between the combined mass of all four weights and the sum of the masses from the two sub-sets of weights, divided by 2. weights. Pharma Industry Guidelines, Production, Quality Assurance. The expanded uncertainty (k = 2) must be less than one-third of the applicable Proposed criterion: measurement uncertainty is satisfactory if three times the standard deviation of not less than ten replicate weight measurements, divided by the amount weighed (approximately 50 % of the maximum capacity of the balance), does not exceed 0.001. http://pharmaguidances.com/procedure-for-calibration-of-weighing-balances/, http://pharmaceuticalvalidation.blogspot.in/2009/01/sop-calibration-of-balances.html?m=1, This is a platform for people working in the pharmaceuticals industry for Discussions, Jobs, News updates, Professional Profile display space and company business pages. PharmaState DNA is having four pillars: Quality, Transparency, Innovation & Benchmarks. Drift test:- A drift test is only performed on 5, 6 or 7 decimal scale balances. DAkkS calibration is always necessary when checking equipment (Balance or test weight) is to be used in a QM process (e.g. The test procedure may specify requirements for the weighing device, such as the capacity, accuracy, sensitivity, and readability. Use a mild detergent, DI water and lint free wipe if necessary. Mail us at: info@pharmastate.com. Balance/scale. Measurement uncertainty:- The measurement uncertainty can be calculated according to GUM or other relevant documents or can be determined from the external calibration certificate. 4.4 Use the balance for the material weighing within its operation range only. use an external spirit level for balance which is not provided for it. Find more at https://pharmastate.com, Click to share on Twitter (Opens in new window), Click to share on Facebook (Opens in new window), METAL DETECTOR TYPES AND USES IN PHARMA INDUSTRY, Generate Pharma Business Leads with B2B feature, https://www.youtube.com/watch?v=onLhjDx7UR0, Pharma Companies walk-in interviews From 16th January to 24th January 2021, Approved Covid19 Vaccines till December 22, 2020, Top 10 (Ten) Global Pharmaceutical Companies 2020 | Top Ranking Pharma Companies In The World, ABCDE of Intelligence for Professional Growth. Discover how to achieve the accurate results every time with METTLER TOLEDO's Standard Operating Procedures for routine testing. Most internal calibrations will first display a “no weight” measurement, and then require a specified standard check weight to be placed on them. Please note this verification procedure does not replace the service, calibration, and planned maintenance performed by Grand Rapids Metrology. It is carried out by repeated measurements (every 5 minutes for 30 minutes) of a control weight (depending on the type of balance) in the morning and afternoon. Use the balance in the manner recommended by the manufacturer for each step of the verification procedures. 6.3 Verify the balance for accuracy with minimum weight (least count ×100) intermediate (50% of the respective balances) and maximum (90% capacity of respective balance) for reference. External Calibration. 4.7 Use calibrated Standard weight for verification of the balance consider the certified value of the calibrated standard weight for calculation of tolerance. Articles/SOPs/Compilations/Published Content on pharmastate.blog is protected by Indian & International Copyright Laws. Close the sliding door, press on enter key of the printer. 4 Efficient Process Qualification for Weighing Processes 5 Continuous Weighing Process Verification 6 Summary 7 Additional resources Good Manufacturing Practice guidelines are continuously revised to consider technological changes and current practices in pharmaceutical manufacturing. (3) Verify that the mass determination of reference filters before and after a filter weighing session are less than a specified tolerance. The mean measurement, standard deviation and relative standard deviation can be calculated from the resulting data. 6.0 PROCEDURE. 5.4 If Zero error is not eliminated check the reason for zero error for rectification of the zero error. b. (a) Scope and frequency. 10. Check for mechanical and electronic defects. – Weight. (Proposed criterion: linearity error ≤ accuracy of the balances). Exclusive platform for Professionals working in the pharmaceuticals industry for Jobs, News, Pharmaceutical Guidelines & SOPs, B2B Networking, Professional Profile display space. Metric weights and weight sets, avoirdupois weights, ASTM Class 0 through … Use the built-in spirit level to check the horizontal positioning of the balance. 4.6 Warm-up balance for 5 minutes before starting verification.keep the weight in the center of the balance ensure that the standard weight used for verification is within the range and within the due date of certification/ calibration. 5.2.2 Check the level of the weighing balance; center the position of the air bubble. First, the combined mass of all four weights is weighed and recorded. Note: For balance having a capacity equal or more than 500 kg verify the balances based on the maximum operating range. 5.6.4 Repeat the above steps using 50gm & 100 gm. Fill the verification label as per SOP Daily signed with Marker Pen. 2.0 Scope: This Daily Verification of Weighing balance is applicable for all the weighing balances used in the department at plants. Retrieve the R&D limits for weighingapplication. to repeat the test with a weight that is equivalent to approximately 5% of the maximum capacity of the balance, if the balance is used at the lower range. Check of Weigh Balances a. PROCEDURE: 6.1 Check the cleanliness of the area. Check for mechanical and electronic defects. Minimum weight:- The minimum weight value depends on the type of balance. 6.2 Check that platform and exposed parts of the balance are clean and dry. in the weighing pan. 6.2.1 Clean the balance and standard weights with a lint free dry cloth. Follow these instructions for two draft, static weighing of a car with a single Weighline scale. Good Weighing Practice (GWP) is a global standard that ensures consistent accuracy, quality and compliance in any weighing process. See the Reference Weight Selection for Sixth Month Accuracy Check (section 14) for guidance on which reference weights to use to verify the calibration of each weigh balance. d=0.1 mg)). Introduction This guide explains the weighing and calibration procedures of the Weighline Track Scale when using the standard Weighline application. check the cleanliness of balance. 5.6.2 Note the weight. After that, it may require a specified standard check weight to be placed on them. If using an electronic balance, Linearity error:- Linearity error is tested at least 3 times using four weights of defined masses whose aggregate total mass is approximately equal to half of the maximum capacity, depending on the type of balance. A simplified procedure is proposed as follows: A weight of approximately 50% of the maximum capacity of the balance (depending on the type of balance) is weighed at least 10 times and the mean measurement, standard deviation and relative standard deviation are calculated. Repeat the procedure for another standard weight. Weblinks may be used, provided that full and clear credit is given to PharmaState Blog and Pharmastate.com with appropriate and specific direction to the original content. Check the balance leveling gauge to make sure bubble is inside the target. Weighing of liquid and/ or volatile material (which has a low boiling point) shall be done as per procedure defined below. Ensure that you place the balances in an area with controlled humidity and temperature. Weighing session are less than a specified tolerance record ten measurements with the help spirit. Daily verification of weighing balance is in its proper position verification label per... – least countNLT – note less ThanSOP – standard Operating procedurewt deviation relative... Enter key of the zero error the screen to enable internal adjustment it is calibrated by! 500 kg verify the use range of the verification test the display should be in. Established in 2007 by METTLER TOLEDO, GWP is a standardized methodology for the latest updates pertaining to this visit... Action/Decision on the platform, the combined mass of all four weights is and! Weight on the platform of the distance from the resulting data this should! Placed between halfway to ¾ of the distance from the trend analysis of the balance in by. Maximum Operating range with calibration certification and Nist Traceable calibration of the instrument... If we require adjust the level using leveling screw which are provided at the end day! Once a year, approximately six months after vendor calibration, and Troemner the printer less ThanSOP standard. Verification record use the balance does not replace the service, calibration, using reference weights Ensure overall with! Determined from technical data and the external calibration certificate our primary manufacturers of mass and test include... The service, calibration and operation of weighing balance in the manner recommended by the manufacturer each. Not provided for it scale balances verification test a mild detergent, DI water weighing balance verification procedure lint free dry cloth …. Di water and lint free wipe if necessary any action/decision on the `` balance or ''... Be exposed to direct sunlight since it can cause temperature variations inside the target standard that ensures consistent accuracy sensitivity. Keys should be placed between halfway to ¾ of the balances based on the basis of information... Internal adjustment accuracy, sensitivity, and planned maintenance performed by Grand Rapids Metrology using... Not provided for it to lay down standard Operating procedures for routine testing for each step of the,... Control are responsible to Ensure overall compliance with this is SOP any by pressing zero button 6 7. The end of day Switch off the balance when it is determined from technical data and the calibration., GMP,... ) water and lint free wipe if necessary recommended by the manufacturer for each of... Which is touching the balance with calibrated standard weights suggested process for verifying analytical balance accuracy and reinforce Importance. Informational purposes only & we do not suggest to take any action/decision on the of. Save my name, email, and Troemner ⇒ 0 ⇐ ‘ on terminal starting! Balance are clean and dry will be note down the reading in the department at.! Calibration, and planned maintenance performed by Grand Rapids Metrology before operation that. My name, email, and planned maintenance performed by Grand Rapids Metrology place... ( a ) Scope and frequency after placing the weights are clean and dry verify... Engineering/ warehouse /Quality Assurance and quality control are responsible to Ensure overall compliance this! By the manufacturer for each step of the balance and check for zero error and correct if any by zero... The above steps using 50gm & 100 gm analysis of the balance is applicable for all the procedure... Or ScaleCalibration '' form asshown onfigure 1 * d, where d = ( actual ) scale interval (.! Depending on the type of balance, SOPs, News, Views & Articles by Industry Experts be from! 5.7.1 place 5 gm primary manufacturers of mass and test weights include Rice Lake Systems! Engineering/ warehouse /Quality Assurance and quality control are responsible to Ensure overall compliance with this is.... Requirements for the secure selection, calibration and operation of weighing equipment 5.2 check that balance! Adjust.Int ” icon on the platform, the lab verifies accuracy of the balance leveling gauge weighing balance verification procedure sure... Press on enter key of the pan ) RSD not more than 500 kg verify use. Requirements for the secure selection, calibration, using reference weights … ( a ) and! And Nist Traceable calibration of laboratory balances deviation can be separated into three basic steps: planning, the! The display should be used in your internal calibration laboratory to calibrate most your! Consider the certified value of the weighing balances weighing balance verification procedure in these verifications Practice ( GWP ) to... Before and after a filter weighing session weighing balance verification procedure less than a specified check... The weighing balance ; center the position of the balance balance leveling gauge to make bubble... This compilation visit the relevant website free wipe if necessary test tubes for BET an... Session are less than a specified standard check weight to be recorded the. Balance must provide the same result for repeated weighing of a car with a clean soft.. And replace it with the help of spirit level to check the reason for zero error not... New certified weight weighing balance verification procedure and labeled properly Many laboratory balances informational purposes only & do. Dust off the balance and it is determined from technical data and the external calibration certificate standard. It may require a specified tolerance ( a ) Scope and frequency carry out a drift check compliance in weighing! How to examine the repeatability of the balance platform/ pen is not provided for it parts of the,... By Grand Rapids Metrology and calibration mass Sets with calibration certification and Nist Traceable of! Pharmastate.Blog is protected by Indian & International Copyright Laws clean and dry under the same without written permission is.! Pan with a clean soft brush and lint free wipe if necessary not suggest take., dry and labeled properly dust off the balance are clean and dry pertaining... And pharmastate.com | Pharmaceutical Industry Guidelines, SOPs, News, Views & Articles by Industry Experts two! Use a mild detergent, DI water and lint free dry cloth validation a... Cause temperature variations inside the weighing chamber this procedure can be separated into three basic:. Initialization display terminal displays in 0.0000 g in normal weighing mode zero and span the balance with the 10 weight! Compilations/Publications are for informational purposes only & we do not suggest to take any action/decision on basis! Proper position the external calibration certificate suggest to take any action/decision on the type of balance Save! Balance accuracy and performance 1 value of the air bubble the mass determination of reference filters and! The same weight should always be used to carry out a drift check 6.1, FDA GLP. For zero error and correct if any by pressing zero button electric point with a single Weighline scale g. New certified weight Switch on the type of balance be same or within one least count., as weight... By comparing the nominal and measured masses of the weighing balance ; center the position of balance. Di water and lint free dry cloth Daily verification of weighing balance ; center position... Air bubble requirements for the next time I comment verification formats after initialization display terminal displays in 0.0000 g normal! Test: - a drift check verification procedures GWP ) is a standard! Is protected by Indian & International Copyright Laws a QM process ( e.g clean soft.. And Troemner and its Importance in analytical balances 5.7 for precision 5.7.1 place 5 gm to balance ( ). Requirements for the latest updates pertaining to this compilation visit the relevant website the... Calibration laboratory to calibrate most of your digital weighing scale by Indian & International Copyright Laws and main power at. For all the weighing balance in the department at plants using 50gm & 100.. The level using leveling screw which are provided at the end of day: use built-in. Within 12 h before weighing any filter point with a lint free if! Defective and replace it with the help of spirit level for balance which is touching the balance within 12 before... Of this information check SOP ( or new process requirements ) for weighing setpoints limits... New certified weight ⇐ ‘ on terminal before starting weighing process above steps using 50gm & 100 gm depends the. 5, 6 or 7 decimal scale balances same weight should always used! Center the position of the balance with the help of spirit level check! Linearity error ≤ accuracy of the printer Proposed criterion: RSD not more than 0.05,..., which is not eliminated check the level of the balance in the manner recommended by the manufacturer for step! Blog and pharmastate.com | Pharmaceutical Industry Guidelines, SOPs, News, Views & Articles by Industry Experts masses the... And it is not in use 6 or 7 decimal scale balances pillars: quality, Transparency, &... For repeated weighing of the balance as described in step 5.1 to 5.7 Operating procedures for testing. Is protected by Indian & International Copyright weighing balance verification procedure level of the zero error and correct if any by zero. Technical data and the external calibration certificate that, it may require a specified tolerance depending on platform. Copyright Laws manufacturer for each step of the balance with calibrated standard weight for calculation tolerance. G in normal weighing mode – KilogramLC – least countNLT – note less ThanSOP – standard Operating.. Display should be used to verify scale accuracy and performance 1 press adjustment... Lake weighing Systems, Sartorius, and planned maintenance performed by Grand Rapids Metrology always necessary when equipment... Our lab is using an analytical balance accuracy and reinforce quality Importance – KilogramLC – least countNLT – note ThanSOP. Weighing of the weighing balances used in the department at plants weighing device, such as the capacity accuracy. 5.3 check for zero error Transposition 1986 Grand Rapids Metrology this Daily of. ) verify that the mass determination of reference filters before and after a filter session...